Technical Information

Measurement and Control Software SGEMCS

Q & A
Q & A index

(Occurred Version)

Q & A list

(Occurred Version)

It may be unable to communicate, if an address is changed in GPIB.(6.0.2)

⇒This problem was corrected in Ver6.0.2.

If "When the *.sgs file loaded,run the communication test automatically" is checked,an error may occur.(6.0.1)

There was fault in Ver6.0.0.

⇒This problem was corrected in Ver6.0.1.

 

If it performs by Windows Vista, I cannot save a settings file. ( - )

Please start SGEMCS by "administrator authority" in Windows Vista. Refer to it here for details.

I cannot install a software in Windows Vista. (3.0.0)

Please use a Windows Vista compatible version (3.0.0 or more). Perform installation by "administrator authority." Refer to it here for details.

At the time of starting of SGEMCS, I get an error message which cannot be read. ( - )

This message is displayed when SGEMCS is started by a general user. A message is not displayed when it is used by a user (an administrator, power user) with the authority more than a power user.

If the password is registered by the user with administrator authority, even if this message is displayed, you can click the "O.K." button and can use SGEMCS.
If you have not registered the password yet, please register it.

The file name read after reading a setting file is not displayed on the output file name of a program instruction tab. (2.0.0)

The problem was in reading a setting file.

⇒It was corrected so that the file name read might be displayed.

Measurement may be well impossible if it re-measures after measurement discontinuation. (2.0.0)

When it measured again after measurement discontinuation, there was a case where the data at the time of the last discontinuation was received according to the timing of measurement discontinuation.

⇒When a stop button was clicked during a measurement and it was in the middle of transmission and reception, it was corrected so that it might stop, after reception was completed.

Can't the test by transmission and reception be performed on the communication test screen of a control devicer? (2.0.0)

The control device could not choose the option button of the method on a communication test window, and was not able to perform it other than "transmission."

⇒It was changed so that all the option buttons of the method could be chosen and reception might also be possible.

Even if the reading wait time is specified on the measuring instrument settings screen, isn't it used at the time of a communication test? (2.0.0)

Even if the reading wait time is specified on the measuring instrument settings screen, it was not used at the time of a communication test.

⇒It was changed so that the reading wait time might be used at the time of the communication test of a measuring instrument.

Can I input the digit count below decimal point into the travel per pulse only to triple figures of a controller setting screen? (2.0.0)

⇒It was changed so that the below decimal point digit count could be inputted to 5 figures.

On the operation screen of a controller, an error might occur at the time of a specified amount of movements. (2.0.0)

When in the case of the controller of SHOT(Mark)-2XX series "999999999" was inputted into the column of the Distance and the move button was clicked, "Relative travel error" occurred.
(It occurred, in order that a controller might not receive "999999999".)

⇒This problem was corrected so that the range of the Distance which can be inputted might be changed from -9999999 by 9999999 and might be checked at the time of an input.

When changing to an Excel instruction tab after reading of a setting file, an error might occur. (2.0.0)

When it changed to the Excel instruction tab after reading the setting file which specified "Correspondence table between columns in Excel and operation devices" as "nothing" on the Excel instruction tab and saved it, the error occurred.

⇒This problem was corrected.

Is the specification method of the output of a result different in Excel instruction measurement and program instruction measurement? (2.0.0)

⇒The specification method of the output of the result in the Excel instruction measurement was changed like the program instruction measurement. Moreover, it was made to measure even if the output place was not specified.

If USB connection of the two controllers is made with PC, one of the two's communication may not be able to be performed. (1.1.3)

The problem was in processing of USB.

⇒This problem was corrected so that both could do communication correctly.

When measuring by a spectroradiometer (SR-3) etc., if a measurement command (commands, such as ST) is used, two or more data will be answered. How should I just receive two or more reply data? (---)

When measuring with a common luminance colorimeter (BM-5 and BM-7) etc., one data will be answered if a measurement command (commands, such as ST) is used. However, the measuring instrument with which two or more data is answered like a spectroradiometer (SR-3) cannot be measured by standard SGEMCS. Since you can use it with a semi- custom-made item, please contact us.

I cannot display the screen of SGEMCS in English. (1.1.2)

SGEMCS can be displayed in English after Version 1.1.2.

I cannot interrupt a program when I specify the number of times as 2 times or more in program instruction tab. (1.1.2)

The problem occurred in auto measure of program instruction. The program has not been interrupted when the number of times was specified to be 2 times or more in program instruction tab. (It does not occur when the number of times is 1 time.)

⇒It corrected so that the problem could be interrupted on the way.

After starting a terminal window, when program instruction measurement was performed, the error occurred. (1.1.2)

The problem occurred in communications processing of program instruction. After starting a terminal window, when program instruction measurement was performed, the error occurred. (It occurs only when performing program measurement after terminal window starting.)

⇒This problem was corrected.

The error might occur at the time of reading of a setting file. (1.1.2)

When the capital letter and small letter of Excel instruction file name or output file name in the setting file differ from an actual file name, the error occurred. (Normally, this error is not occurred unless the capital letter of a file name and a small letter are changed after writing setting file.)

⇒This problem was corrected.

When using a controller by closed loop control method in Excel instructions measurement, a stage may be unable to be moved to the specified coordinates value. (1.1.2)

When 'close' was set as the control method on the setting window of a controller (MS series), the error occurred in stage movement of Excel instructions measurement.
(Only when using a controller by closed loop control method in Excel instructions measurement, the error occurs.)

⇒This problem was corrected.

When I click 'x' button at the upper right of the axis control window and close it, I cannot display the window again. (1.1.1)

The problem occurred in display processing of axis control window. When x button at the upper right of the axis control window is clicked and it is closed, the window cannot be displayed again. (When exit button of the axis control window is clicked and it is closed, the problem does not occur.)

⇒This problem was corrected so that the window can be displayed again when x button at the upper right of the axis control window is clicked and it is closed.

When I input 0 into the move pitch in the controller settings window, I get an error message telling me to "having divided by 0". (1.1.1)

The problem was in processing of a move pitch input. When 0 was inputted into the move pitch in the controller settings window, the error of "having divided by 0" was displayed. (Usually, since a pitch is not specified to be 0, the problem does not occur.)

⇒This problem was corrected so that an error might not occur, even if it inputted 0 into a move pitch.

In the axis control window, the 2nd axis of MINI-5D has also been set as the same speed as the 1st axis. (1.1.0)

The problem was in speed settings of MINI-5D. In the axis control window, when a move button is clicked, the 2nd axis has also been set as the same speed as the 1st axis.

⇒This problem was corrected so that speed might be set up as a setting value by the 1st axis and each 2nd axis.

A stage may not move as the specified amount of movements. (1.1.0)

The problem was in calculation of the amount of movements of a stage. When the amount of movements was a decimal value, the error arose by the calculation of a decimal and there was a case where exact movement could not be performed.

⇒It corrected so that an error might not arise in calculation.

At the time of the measurement start of program instruction and Excel instructions measurement, the speed of a controller was set up only for the 1st axis. (1.1.0)

The problem was in speed settings of program instruction. At the time of the measurement start of program instruction and Excel instructions measurement, the speed of a controller was set up only for the 1st axis.

⇒This problem was corrected so that the speed of each axis might be set up correctly.

By program instruction, when a move pitch was larger than 32767, and when move pitches were -1 and -2, it did not measure correctly. (1.1.0)

The problem was in stage move processing of program instruction. By program instruction, when a move pitch was larger than 32767, and when move pitches were -1 and -2, it did not measure correctly.

⇒This problem was corrected so that it might measure correctly.

Functional Addition / Specification Change

(Corresponding Version)

    • Setting file (an extension is .sgs) was related with SGEMCS. (4.0.1)
    • It corresponded to license management of the software by the USB key. (4.0.0)
    • It corresponded to WindowsVista. (3.0.0)
    • The graphic type dialog was added to the Axis Control dialog. (2.0.6)
    • In program instruction measurement, the measurement method which moves two or more axes of a stage simultaneously was added. (2.0.6)
    • It corresponded to FINE-01. (2.0.6)
    • "Save or Load Device File" menu was added to the File menu. (2.0.6)
    • In excel instruction measurement, "STOP" which is a command for a stop during measurement was added to the command described to an instruction file. (2.0.6)
    • Even if the communication error occurred during measurement, it was made not to display a "communication error" message, when retry is set as the measuring instrument to be used by program instruction measurement. (2.0.5)
    • The license management method of software was changed.
      Anyone can use it for 30 days after software installation.
      To continue using 30 days after, the key (password) to open the lock for license management needs to be registered.
      Please refer to "About the license of software.pdf" about a registration method. (2.0.4)
    • When "Return to start position after measuring" was chosen by Excel instruction measurement, it could be chosen whether a starting position would be cleared to zero with an instruction sheet at the time of "R" command execution. (2.0.3)
    • It corresponded to SHOT-302GS and SHOT-304GS. (2.0.3)
    • When "Process all instruction sheets" was chosen by Excel instruction measurement, it could be set up whether the check message before the measurement start of the sheet after the 2nd sheet would be displayed. (2.0.2)
    • It corresponded to FINE-501, 502, and 503. (2.0.2)
    • It corresponded to OMEC-2BF, 2BG, 4BF, and 4BG. (2.0.1)
    • When a controller operation setting screen was opened, a setup of all axes was displayed. (2.0.0)
    • In the case of the controller, the name input column was added to the device settings screen. (2.0.0)
    • The automatic measurement in program instruction measurement can be performed now combining two or more controllers. (2.0.0)
    • The angle was added to the unit of a controller settings screen.
      (Except MINI-5D and Mark102) (2.0.0)

The instruction file creation helper program (Ver.1.0.0) was created. (2.0.0)

  • It corresponded to Mark-501, Mark-502, and PAT-001. (2.0.0)
  • It can be specified now as the measurement column of an Excel instruction file combining number specification (#1) of a measuring instrument, and wait specification. (2.0.0)
  • It came to be able to perform discontinuation and resumption during Excel instruction measurement (for it to usually measure) execution. (2.0.0)
  • The sign display of "+" and "-" was added to the move button of an operation screen. (1.1.3)
  • The button for rerun of a current line was added to the step measurement screen of Excel instruction measurement. (1.1.3)
  • Rerun of a current line can be chosen now at the time of the error message display under Excel instruction measurement. (1.1.3)
  • The initial value of magnescale value in the controller (MS series) settings window was changed into 100. (1.1.2)
  • The model name of a controller was changed and it corresponded to SHOT series. (1.1.2)
  • The specification of devide character was added in measuring instrument settings window. And characters other than a comma can also divide the measurement value. (1.1.1)
  • The specification of read wait was added in measuring instrument settings window. And waiting time can be put in between transmission of a measuring instrument, and reception. (1.1.1)
  • It was surely made to perform one line at a time, without omitting the line specified "no measurement." at step measurement of Excel instruction.
    Even if not measured, the instructions of a controller or a control device were written in an output sheet.
    Even if not measured, it enabled it to click a mark button. (1.1.1)
  • When an error message was displayed by Excel instructions measurement, the device name was displayed on the error message. (1.1.1)
  • The shortcut key was assigned to the button of a step measurement window of Excel instruction. (1.1.0)
  • The check box which "Display error messages" was added to Excel instruction window. (1.1.0)